Gantry Software Updates
7-10
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
not, continue with the instructions below.
After upgrading the EDC board’s micro con-
troller firmware, the ECG connection must be
restored and the EDC module must be power
cycled:
14.
Remove the serial interface cable from the
connector A1 RS232 and reconnect the
ECG cable (see step 2 in the section
Preparing for Firmware Upgrade
on page
7-8).
15.
Remove the EDC power cable connected
to connector A4 POWER. Reconnect the
power cable.
Proceed to upgrade the micro controller firm-
ware of the other EDC board:
16.
Connect the serial interface cable to the
other EDC board as described on page 7-8.
17.
Repeat steps 1 (page 7-8) to 15 (page 7-
10) of this procedure to update the micro
controller firmware of the EDC module on
the other detector.
18.
When both EDC modules have been up-
dated, remount the detector top cover fol-
lowing instructions provided in Chapter 6.
Repair Procedures
.
Upgrading Logic Firmware
Procedure Details
To download logic firmware to the EDC
board, you use the terminal emulator program
ttermpro. Proceed as follows:
1.
In the Windows Explorer, select the folder
EDC CardioMD programming for
Module rev. 9CSY0900-B02 or
later\TTERMPRO
on the CD-ROM.
2.
In this folder, double-click the icon
ttermpro.exe
to start the ttermpro pro-
gram.
Connector A4 POWER
on the EDC board
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......