Setting Up FTP and Telnet
5-54
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
5.
Enter the new value
0
[Enter].
This setting disables NTLM.
6.
When prompted, enter
y
[Enter] to
confirm.
7.
Enter
0
[Enter] to exit the menu.
8.
Enter
0
[Enter] to quit the application.
9.
In the Administrative Tools window,
double-click Services.
10.
In the right-hand pane of the Services
window, double-click Telnet.