Detector Calibration Files
5-58
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
> CardioMD
[Enter].
The acquisition PC answers: Welcome to
Microsoft Telnet Server. You can now enter
DOS commands to operate the acquisition PC.
5.10
Detector Calibration Files
The directory
C:/Cardiocam/Calibrations
stores five calibration files per detector. The
files are in ASCII format and can be reviewed in Windows Notepad. The five files are:
•
dyncor.det#
•
Ecal.det#
•
Geocal.det#
•
HVcal.dat
•
PMTcal.det#
where
#
is
1
for detector 1 and
2
for detector 2.
Figure 5.16 on page 5-59 shows typical values in the files together with a brief explanation.