Diagnostics
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
5-33
Command
Rotate controller
Linear controller
RA
Query the Runaway status for the actual motion controller. Valid for both
motion controllers as each one individually may activate a Runaway output
which removes motor power.
If Runaway is not active, the response will be:
No runaway, mask = 0x8467
(the value for ‘mask’ differs for the two motion controllers and may also
differ for different software versions).
If Runaway is active, the response will comprise a cause code:
RUNAWAY, cause code = 0x0020, mask = 0x8467
The cause code is bitmapped, i.e. there may be more than one cause present
at the same time. The interpretations of the individual bits are:
RA_X_ENC_ERROR
0x0001
The position readings from the
absolute and the incremental
encoders for the X axis differ too
much
RA_Y_ENC_ERROR
0x0002
Same for the Y axis
RA_Z_ENC_ERROR
0x0004
Same for the Z axis
RA_LOST_SYNC_
ERROR
0x0020 Checksum
error on 3 consecutive
messages on the serial link used for
communication between the two
motion controllers. Happens if one
of the motion controllers stops
responding
RA_TIMEOUT
0x0040
The linear controller has not
received synchronization messages
from the rotate controller at the
required regular interval
RA_X_ILL_MOVE
0x0100
The X axis has moved when it is not
expected to
RA_Y_ILL_MOVE
0x0200
Same for the Y axis
RA_Z_ILL_MOVE
0x0400
Same for the Z axis
RA_TRAP
0x4000
The micro controller has encounterd
an illegal instruction
VERSION
Print the boot message including software version (see the section
Boot
prompt
on page 5-28)
SSCTX
Transmit a message on the high-speed serial link used for communication
between the two controllers.
Do not use unless you receive further instructions
BAUD
Change the baud rate of the interface. Remember to change the terminal
baud rate immediately afterwards
HELP, ?
Print list of commands
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......