CardioMD AC Option
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-75
12.5.3.9
Opening the Line Source Attenuator
After approximately one year of using a line source, the line source housing’s attenuator will have
to be opened in order to ensure that the line source is used optimally.
Procedure Details
1.
Remove the line source housing from the
line source compartment, as described in
the section
Removing the Line Source
Housing
, page 12-68.
WARNING
Make sure that the transportation screws are in place and locking the shutter
before attempting to remove the source holder. Failing to lock the shutter may
cause you to be exposed to radiation from the source inside the holder.
Procedure Details
The photos on the right show the attenuator in
the closed position. The attenuator is in this
position when the line source housing is
delivered.
CAUTION
RADIOACTIVE
MATERIAL