Repair Procedures
12-68
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
12.5.3.8
Exchanging Line Sources
The line sources of the AC option must be replaced approximately every two years.
WARNING
When exchanging line sources, do
not
open the line source housing to remove
the line source. The entire line source housing must be returned to the
manufacturer for exchange of the source.
Removing the Line Source Housing
Before the line source housing can be removed from inside the line source compartment, you must
dismount the transmission scanner from the detector, detach the line source compartment from the
AC option bracket and remove the covers of the line source compartment. Proceed as follows:
Procedure Details
1.
Remove the transmission scanner from the
detector.
See section 12.5.1.1, page 12-29 for
instructions.
2.
Detach the line source compartment from
the AC option bracket.
See section 12.5.1.2, page 12-33 for
instructions.
3.
Remove the covers from the line source
compartment.
See section 12.5.3.1, page 12-38 for
instructions.
CAUTION
RADIOACTIVE
MATERIAL
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......