Gantry Software Updates
7-12
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
11.
When both EDC modules have been up-
dated, remount the detector top cover fol-
lowing instructions provided in Chapter 6.
Repair Procedures
.
7.2.4
Collimator ID Programming
The collimator type ID is programmed into an EEPROM located on the collimator contact board.
7.2.4.1
Tools Required
•
CardioMD software installation CD, part number 9CSY0906
•
NULL modem serial interface cable (DB9-female to DB9-female).
•
When using a laptop acquisition PC (CardioMD Series III): An USB to RS 232 adapter, part
no. 3ACQ1669.
7.2.4.2
Procedure
Before starting the collimator ID programming, check the release note on the CD-ROM. You find
the release note in the directory
Release notes
. The release note may provide information that
is more recent than this manual.
Procedure Details
1.
Remove the detector top cover to access
the EDC modules. See instructions in
Chapter 6
Repair Procedures
.
2.
Remove the ECG cable from the ECG port
labeled A1 RS232.
3.
Connect one end of the serial interface
cable to this port.
If you are using a laptop acquisition PC:
4.
Use the USB to RS 232 adapter to connect
the opposite end of the serial interface ca-
ble to one of the acquisition PC’s USB
ports.
See section 7.2.2, page 7-2 for instructions on
determining which COM port you are using.
If you are using a tower acquisition PC:
5.
Connect the opposite end of the cable to
the acquisition PC’s COM1 serial port.
III
I, II
Serial interface
cable
ECG cable
III
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......