Mounting and Dismounting the LSM 710 / LSM 780 Scan Head
The scan head is connected to the microscope via an integrated safety interlock. It can be moved
between two microscopes. Make sure the system is shut off completely before starting the following
procedure:
Be aware that the scan head weights up to 28 kg.
Please see chapter 1 for safety relevant cables and connections.
Moving the scan heads between Axio Observer.Z1, Axio Examiner and Axio Imager.Z2:
•
Loosen the three fastening screws of the LSM 710 / LSM 780 (Fig. 6/
1
and Fig. 7/
1
).
•
Slowly pull the scanning module (Fig. 7/
2
) away from the microscope port or the tube (Fig. 7/
3
).
•
For mounting the scan head onto a microscope, make sure the pins and the electronic connections of
the safety interface match closely.
•
Push the scan head to the microscope or the tube.
•
Fasten the three fastening screws of the LSM 710 / LSM 780 (Fig. 6/
1
and Fig. 7/
1
).