MACROS AND VISUAL BASIC
LSM 710 and LSM 780
Carl Zeiss
Macros
Systems
16 M60-1-0025
e
02/2010
•
In order to follow the calibration process, just
simply click into the Message Window and; a
scroll bar will appear (Fig. 16).
•
The Calibration Matrix takes about 25 min for
each objective.
•
After the macro has finished, a message will
pop up with the request to switch the objective
lens in order to make the changes effective.
•
If necessary, repeat the procedure for the other
objectives.
•
Use the
Update Diagrams
button in order to
check, if an objective has already been cali-
brated.
−
Select an objective, press the
Update
Diagrams
button
−
If no graphs appear, this particular objective
has yet needs to be calibrated.
•
Click
Exit
to finish.
Meaning of the Graph (Figure):
−
Each coloured line represents an individual laser line.
−
Each coloured calibration point represents an emission filter.
−
Standard graphs are displayed with black points.
Meaning of the Colours:
−
<420 nm:
Æ
Pink
−
421…475 nm:
Æ
Blue
−
476…495 nm:
Æ
Turquoise
−
496…550 nm:
Æ
Green
−
551…590 nm:
Æ
Yellow
−
>590 nm:
Æ
Red
Fig.
16
Collimator Calibration window after
calibration