NOTES ON DEVICE SAFETY
Systems
User Interface
Carl Zeiss
02/2010 M60-1-0025
e
31
1.11.3
Mounting and Dismounting the Scan Heads
The scan heads are connected to the microscope via an integrated safety interlock. They can be moved
between two microscopes. Make sure the system is shut off completely before starting the following
procedure.
Be aware that the scan head weights up to 28 kg.
•
Depending on the microscope stand, it might be necessary to remove the HBO lamp house first
before removing the scan head.
•
In case of the Axio Imager stand with an angled adapter on the HBO lamp house, the adapter will
remain on the stand while the lamp house is removed!
Moving the scan heads between Axio Observer.Z1, Axio Examiner and Axio Imager.Z2:
•
Loosen the three fastening screws of the LSM 710 / LSM 780 (Fig. 24/
1
and Fig. 25/
1
).
•
Slowly pull the scanning module (Fig. 25/
2
) away from the microscope port or the tube (Fig. 25/
3
).
•
For mounting the scan head onto a microscope, make sure the pins and the electronic connections of
the safety interface match closely.
•
Push the scan head to the microscope or the tube.
•
Fasten the three fastening screws of the LSM 710 / LSM 780 (Fig. 24/
1
and Fig. 25/
1
).
Fig. 24
Port connection between LSM and
Axio Observer.Z1
Fig. 25
Port connection between LSM and
Axio Examiner or Axio Imager.Z2