LEFT TOOL AREA AND HARDWARE CONTROL TOOLS
LSM 710 and LSM 780
Carl Zeiss
Acquisition Tab
Systems
54 M60-1-0025
e
02/2010
Dye
(Fig. 41/
2
):
•
Once finished with the input,
Smart Setup
suggests alternative considerations: One for
fastest
imaging, one for the
best signal
,
best compromise
between both speed and best signal and the
optimal setup for later
linear unmixing
of the dyes.
•
Clicking
Apply
, automatically sets the ideal hardware parameters for the dyes chosen. If the option
Linear Unmixing
is chosen, the system is set in the lambda mode automatically. Clicking the
Auto
Exposure
button will optimize the settings of the Gain (Master) and offset for the given
laser power and pinhole size. Further image optimization from this point can be done easily.
Fig. 41
Smart Setup menu
For questions or suggestions send an e-mail to
.