LSM 710 and LSM 780
CENTER SCREEN AREA / IMAGE CONTAINERS - …
Systems
Raster Scanning Image Correlation Spectroscopy
Carl Zeiss
02/2010 M60-1-0025
e
251
When
Immobile structures (+Stack)
is selected, the average image from the stack (pixel by pixel) is
subtracted from every image and a scalar is added to avoid negative pixels. The scalar is computed from
the average of all pixels from all frames of the stack. When
Immobile structures (+Plane)
is selected,
the average image from the stack (pixel by pixel) is subtracted from every image and a scalar is added to
avoid negative pixels. The scalar is computed from the frame to which it is added. Both procedures are
equally valid in most of applications.
6.18.4
Display Functionalities
Activate
Display
under the
RICS
view. The Display options will be displayed in the View control window
(Fig. 286).
Fig. 286
Display control window
In the
Representation
pull down menu you can
choose how the correlation function should be
displayed.
Four options are available (Fig. 287):
Diagram
will display the horizontal and vertical
profiles of the correlation function and the fit /
residuals if activated;
Image
will display the correlation and the fit /
residuals if activated as two dimensional images;
Table
will display the coordinates of the horizontal
and vertical profiles of the correlation and the fit /
residuals if activated that can be copied or saved
via a right mouse click from the
Table Export
menu (Fig. 288);
2.5 D
will display the correlation function and the
fit / residuals if activated as pseudo 3D images
(Fig. 289).
Fig. 287
Representation pull down menu
Fig. 288
Table export menu