NOTES ON DEVICE SAFETY
Systems
Notes on Care, Maintenance and Service
Carl Zeiss
02/2010 M60-1-0025
e
27
•
Unlock the lamp housing (Fig. 18/
1
) by slightly turning it anticlockwise.
•
Remove the halogen lamp with alignment base (Fig. 18/
3
).
•
Remove new 12 V 35 W halogen lamp with adjustment base from the packing box. Insert it in the
mount taking care that the tip of the mount engages in the centering notch of carrier plate.
•
After replacing the halogen lamp, attach the lamp housing and lock it.
Fig. 18
Replacing the halogen lamp
Care operations that may be carried out by operating staff are limited to cleaning painted and glass
surfaces.
•
Before cleaning the instrument make sure the main power supply is disconnected.
•
Cleaning painted surfaces:
To do this, use a clean cloth that has been moistened in a mixture of water and some detergent; do
not use any solvent, however. Dry with a lint-free cloth.
•
Cleaning glass surfaces:
Glass surfaces that have become soiled or which are marked with fingerprints may be rubbed with a
clean optical cleaning cloth.
If soiling is persistent, dip the optical cleaning cloth into a mixture of distilled water and a small
quantity of detergent.
To complete cleaning, lightly breathe on the glass surface and rub it dry with a clean cloth. Lint or dust
is best removed with a clean brush.
•
Make sure that no cleaning liquid penetrates into the system.
We strongly recommend to get the calibration objective for the objective turret, since the calibration and
test tasks in the system maintenance tool (see CHAPTER 6: "TOOLS, ADDITIONAL SOFTWARE") require
this tool.
Especially the scanner calibration is recommended every 1-2 years, when the high speeds (13-15) are
used frequently.