CONFOCOR
3
ConfoCor 3
Options
Carl Zeiss
02/2010 M60-1-0025
e
133
To this end use an edge of a structure or a line. If
the
Scan Corr
is not adjusted, you will see a pixel
shift (Fig. 148). Alter the
Scan Corr
values to
compensate any pixel shift in the following way:
Move the
Scan Corr X
slider until the pixel shift is
at the minimum (Fig. 149). Rotate the image by
90° and redo the same adjustment for
Scan
Corr Y
.
9.13.4
Auto Save
The
Auto Save
panel (Fig. 150) enables you to
instruct the program to save the results
automatically and continuously during the
measurement process. All data recorded until a for
example a software failure will be available this
way. This function is recommended when
performing long measurements. If the
Auto Save
check box is ticked you can specify the directory in
the
Directory
selection box, to which the
measurement should be saved.
Only .fcs and hence ANSI format files are saved in this directory. .raw data files are still saved in the
directory specified under
Save raw data during measurement
. However, if the
Auto Save
option is active, temporary raw data files will be saved permanently and not only temporarily and
will be available if a software failure occurs. The
FileName
entry indicates that .fcs files will be
saved. This entry cannot be changed.
Fig. 148
Image of a line before pixel shift
compensation in x
Fig. 149
Image of a line after pixel shift
compensation in x
Fig. 150
Option for ConfoCor, Auto Save
window