CONFOCOR
3
Carl Zeiss
Acquisition
ConfoCor 3
40 M60-1-0025
e
02/2010
If well positions are changed the system will automatically activate the
XY Stage
tab. The
X,Y
Stage
tab will only be available for scanning stages. It is deactivated for manual stages. Even if the
Scanner
button is activated, selecting a well will automatically deactivate the
Scanner
and
activates
the X,Y-Stage
mode. If you take an image of the well than the position can be changed
in
Scanner
mode within the well.
When the required chamber is approached by joy stick or the compass rose, orientation of the
carrier will be lost. Thus positioning has to be done visually via inspection of the carrier in this case.
If the
XY Stage
button is activated, all positioning will be done via the scanning stage.
This mode corresponds to the
Sample Carrier
mode, if
Single Position
is selected. Regardless
which well is selected, in the result table of the
FCS results
window the position is indicated as
"
1
"
.
Scanner control
•
Activate the
Scanner
control tab to activate scanning mirror positioning.
•
Move the scanning mirrors in the appropriate position by clicking on the arrow keys of the compass
rose. Each mouse click moves the scanners in the appropriate direction by one step.
•
Set the required step width via the
Step-size (µm)
the input box or by using its scroll arrows. The
travel speed of the microscope stage can be set to steps
1
,
2
and
3
. Step 2 is preferably used for
standard positioning.
•
The current x and y positions are displayed in the
x-Position
and
y-Position
input boxes. You can use
the input boxes to change the position by typing in a number or using the scroll arrows. The next click
on one of the compass rose arrows will bring the laser beam via the scanning mirrors r to that
position. The following click will move the scanning mirrors with the current speed and step sizes in
the indicated direction.
If the
Scanner
button is activated, all positioning will be done via the scanning mirrors using the
compass rose. However, using the joystick or clicking on a well will automatically activate the
XY Stage
tab and movement will be done by the stage.
There are three ways to position the laser beam with respect to the sample by pressing either of the
following buttons:
−
Current Position:
no automated positioning possible. Sample has to be manually positioned in
respect to the laser beam
−
Sample Carrier:
automated positioning used for maneuvering between wells of a carrier
−
LSM image:
automated positioning
using an LSM image