LSM 710 and LSM 780
LEFT TOOL AREA AND HARDWARE CONTROL TOOLS
Systems
Acquisition Tab
Carl Zeiss
02/2010 M60-1-0025
e
65
QUASAR detector settings
Use the two sliders to define the spectral detection range of the QUASAR detector.
Resolution
: Step size can be 3.2, 4.9, 9.7, 19.4, 23.3 or 38.9 nm (slightly different values for
GaAsP QUASAR).
5.2.5.8
Online Fingerprinting Mode
(to be selected in the
Imaging Setup
tool).
The use of this function permits the selection of
reference spectra together with the excitation
settings, allowing an immediate display of the
unmixed images while scanning. Chose
Online
Fingerprinting
in the
Imaging
Setup
tool to
activate the according display in the
Light Path
tool.
You can change the settings using the function
elements which have been described for the
Lambda Mode
plus the following:
RS1 ... 8
Menu for selecting a display color and reference spectra
(reference spectra derive from earlier experiments via
Mean of ROI functionality or ACE tool). The unmixed
results are displayed during scanning and Lambda Stack
data will neither be displayed nor stored.
Fig. 55
Light Path tool window;
Online Fingerprinting Mode