LEFT TOOL AREA AND HARDWARE CONTROL TOOLS
Systems
Ocular Tab
Carl Zeiss
02/2010 M60-1-0025
e
51
(3)
Microscope settings on Axio Observer for reflected-light observation (Epi-fluorescence)
•
Turn on the HBO 100 power supply switch (Fig. 35/
1
).
•
Click on the
Reflected Light
button and set the shutter to
Open
.
•
Click on the
Reflector
button and select the desired filter set by clicking on it.
•
The filter is automatically moved into the beam path to enable observation in epi-fluorescence.
•
Click on the
Objective
button and select the objective.
•
Make sure that the illumination from the transmitted-light is switched off.
5.1.1.3
Controls for Axio Examiner
For setting the Axio Examiner, proceed in the same way as with Axio Imager.Z2 and Axio Observer.Z1.
Since the objective turret of the Axio Examiner is not motorized the setting of the desired objective has to
be done manually.
The objective in use must be indicated in the
Acquisition Mode
tool.
5.1.1.4
Controls for the Axio Scope with Motorized Z-drive
The Axio Scope (only available for LSM 700) has only a motorized z-drive, all other components are not
encoded and have to be set in the software manually for correct functionality. Especially, the objectives
are changed manually. The used objective must be set in the
Acquisition Mode
tool.
5.1.2
Ocular Tool – Incubator Control
From ZEN 2010 the incubation settings can be set
within the main software.
The settings for Temperature (Ch1 and Ch2) and
CO2 can be set by using the input boxes.
The measured values are displayed next to the
input boxes.
Fig. 36
Ocular tool, incubator control