Maintenance Tool
Systems
66 M60-1-0025
e
02/2010
Scanner linearity test (Scan field):
=> Test tool: calibration objective (turret will be set automatically)
Subset of task scan field checks the scanner linearity and runs much faster than scanner calibration.
Resolution test
=> Test tool: calibration objective (turret will be set automatically)
Checks the resolution of the fine structures on the calibration tool. If the resolution test is not successful,
this hints on wrong collimator (and eventually pinhole) positions.
Illumination (optional):
Illumination test (fluorescence) for homogeneous coverage
=> Test tool: Delta Vision Slide orange.
After each call the image exposure needs to be set, so that no magenta-colored pixels can be seen.