TOOLS, ADDITIONAL SOFTWARE
LSM 710 and LSM 780
Carl Zeiss
3D for LSM
Systems
58 M60-1-0025
e
02/2010
Automatic Object Measurement - General
This function carries out an automatic measurement and labeling.
Measured
Object Features
Measured
Volume Features
Fig. 40
The regions must be defined by an image sequence
Mask Image
(the objects must be separated from
one another by black voxels with the grey value 0). This sequence is generated with the function
Segment
. If it is a multichannel sequence a single channel has to be chosen.
The image
Dens Image
is needed for the measurement of the densitometric features. Image sequence
properties like scaling and unit are taken from
Dens Image
. A single channel of this sequence (if it is
multichannel) must be selected with the buttons to the right of the parameter.
The measurement results can be stored to database files. These files are tab delimited ASCII files which
can be easily imported to major Windows programs like text processing or spreat sheet application.
Writing database files are independently supported for object and volume features. Activating the
corresponding check boxes enables it. The name of the database is defined with the field
Database
. The
files will be located in the subdirectory DATA of the main installation directory. The filename extension
TXT will be added automatically.
If the check box
Label
is activated a single channel sequence will be generated. It contains all the
measured objects, each object is coloured homogeneous but in different colours. To copy all
measurement values to the clipboard activate the check box
Clipboard
.