LSM 710 and LSM 780
LEFT TOOL AREA AND HARDWARE CONTROL TOOLS
Systems
Ocular Tab
Carl Zeiss
02/2010 M60-1-0025
e
45
(1)
Microscope settings on Axio Imager for
transmitted-light
observation
•
Set the reflector turret position to
None
and
click the
On
button for transmitted light.
•
Pull out the push rod (Fig. 33/
1
) to switch the
beam path of the tube in LSM position.
•
Control the brightness of the halogen lamp
with the potentiometer (Fig. 33/
4
) or the
Intensity %
slider in the
Ocular
panel.
•
Set the required transmission value of the gray
filters in the
Filter
frame.
•
Set the condenser and the luminous-field
diaphragm for KÖHLER illumination.
The laser-scanning mode automatically occludes
the halogen lamp even with
Transmitted Light
activated (
On
).
•
Click on the
Ocular
tab in the Left tool area.
Press
Online
if needed.
•
Place specimen on microscope stage.
−
The cover slip must be facing up.
Fig. 32
Axio Imager.Z2 controls in the Ocular
panel