SETUP REQUIREMENTS
LSM 710 and LSM 780
Carl Zeiss
Scanning Module LSM 710, LSM 780, LSM 710 NLO and LSM 780 NLO
Systems
8 M60-1-0025
e
02/2010
2.8
Scanning Module LSM 710, LSM 780, LSM 710 NLO and LSM 780 NLO
Scanners
2 individually driven galvanometric scanners
Scanning speed
Up to 8 frames/sec (512
×
512 pixels)
Field resolution
Max. 6144
×
6144 pixels (individually adjustable for each axis)
Field of view
11.3
×
11.3 mm² with a 1.25
×
objective (20 mm diagonal in intermediate
image)
Zoom
0.6
×
... 40
×
, continuous control (Axio Examiner 0.67x to 40x)
Channels
a) Up to 3 confocal fluorescence channels (PMT) simultaneously
or
b) 34 confocal fluorescence channels (PMT-array) simultaneously
or
c) 32 GaAsP + 2 PMT fluorescence channels simultaneously (LSM 780)
1 transmitted light channel (PMT) and up to 5 NDD channels
or
2 - 10 NDD channels (Non descanned detectors, PMT)
2 monitor diodes for maintenance purposes
Direct adaptation of external detectors
Dynamic range
16-bit DAC for each detection channel
Pinholes
1 pre-adjusted master–pinhole, individually adjustable for Multitracking and
short wavelengths (i. e. 405 nm)
Computer controlled automatic alignment
2.9
Detection Module ConfoCor 3
Channels
2 channels with APD detectors
Dynamic range
12-bit DAC for each detection channel
Pinholes
1 individual pinhole (channel 4 of LSM 710, LSM 780, LSM 710 NLO and
LSM 780 NLO)
Computer controlled automatic alignment
2.10
Detection Module LSM
BiG
Channels
2 channels with GaAsP detectors
Dynamic range
8-, 12- or 16-bit
Pinhole
Master pinhole of the LSM 710 with automatic alignment