LSM 710 and LSM 780
STARTUP AND SHUTDOWN OF THE SYSTEM
Systems
Startup of the System
Carl Zeiss
02/2010 M60-1-0025
e
15
Maintenance Tool
icon
With a calibration objective and correct system configuration, the maintenance tool
allows convenient self adjustment of the LSM 710 and LSM 780 systems system. The
optical beampath, relative pinhole position and scanner adjustment can be set and
checked automatically on the LSM 710.
Stand Select
Stand Select
icon
The
Stand Select
tool permits a new or updated database to be assigned to the
LSM software program. This function should preferably be performed by authorized
service personnel. See printed manual CHAPTER 6 TOOLS, ADDITIONAL SOFTWARE.
ZEN
icon
Start ZEN software for operating the LSM laser scanning microscope.
3.1.1
Handling of the Argon Laser
1
Holding-down
bolt
6
Power measuring connection
2
Connection to the main power supply unit
7
Connection for external interlock loop
3
Red LED (output power too high -
8
Switch LASER RUN/STANDBY
strong reduction of lifetime)
9
Green LED (Laser power stabilized; optimum power
4
Light control potentiometer (laser power)
range regarding lifetime)
5
Current measuring connection
Fig. 3
AR laser control