STARTUP AND SHUTDOWN OF THE SYSTEM
LSM 710 and LSM 780
Carl Zeiss
Shutdown Procedure
Systems
18 M60-1-0025
e
02/2010
3.2
Shutdown Procedure
Never shut down the computer by its main switch while the ZEN software is still active, or else
you will lose the currently set operating parameters and the images just scanned.
In the
Maintain
tab /
LSM Options
dialog window, activate
Laser off
on
Exit
in the
Shutdown
tab (section
LSM Options
). The lasers will then automatically be switched off when
you exit the LSM program.
3.2.1
Exiting ZEN Software
•
Make sure you have saved all your image data.
•
Switch of all the lasers which might still be running.
•
Close the ZEN software window. This terminates the ZEN software.
•
Shut down WINDOWS VISTA
•
About 20 seconds after WINDOWS VISTA is shut down your computer turns off.
3.2.2
Switching System Power Off
Please bear in mind that a cooling phase of at least 5 minutes is required between switching off
of the lasers and switching off of the entire system via the power remote switch or the Power
Supply switch.
5 minutes after computer shutdown
set the power remote switches
System/PC
and
Components
or the
Main
switch on the System Electronic Rack to position
OFF
.
This puts your LSM 710 / LSM 780 microscope system, including the computer, off power.