SETUP REQUIREMENTS
Systems Vibrations
Carl
Zeiss
02/2010 M60-1-0025
e
7
2.6
Vibrations
Vibrations under operation conditions
(with system table)
Shipping shock (LSM 710 / LSM 780 box)
5 µm pp at 5 Hz
10 g
10 µm pp at 10 Hz
10 µm pp at 20 Hz
2.7
Microscopes
Inverted Axio Observer.Z1 with side port or rear port
Upright Axio Examiner with LSM-tube or rear port
Upright Axio Imager.Z2 with LSM-tube or rear port
Upright Axio Imager.M2p with LSM-tube
All IC
2
S objectives from Carl Zeiss and their accessories can be
accommodated.
Z motor
DC servomotor, opto-electronically coded
Least Z interval:
25 nm
(Axio Observer.Z1 SP or RP)
30 nm
(Axio Examiner)
10 nm
(Axio Imager.Z2)
25 nm
(Axio Imager.M2p)
measured according to VDI 3441
Piezo Objective focus
Piezo-driven single objective drive
Max. travel 250 µm; resolution 15 nm
Z-Piezo stage
Piezo driven stage insert for scanning stages
Max. travel 250 µm; resolution 5 nm
In the unlikely case of extreme fluctuations of the external power net or
electromagnetic radiation, the piezo crystal will vary and disturbance in
the image is visible. Note that this is not a defect and the piezo drive will
not be damaged.