INTRODUCTION TO THE SOFTWARE APPLICATION LAYOUT LSM 710 and LSM 780
Carl Zeiss
Menu Bar
Systems
28 M60-1-0025
e
02/2010
4.5.2
Maintain
The
Maintain
menu from the
Menu
bar contains
additional modules to check and guarantee the
interference-free operation of all the software and
hardware components of the LSM 710 and LSM
780 systems. In the
Menu
bar, click on
Maintain
.
This opens the Maintain list (Fig. 12).
The maintain list is bipartite. On upper section lists
user accessible tools, the lower section lists tools
reserved for service personnel, they are password
protected.
See also section
Maintain Tab
for further
information.
4.5.2.1
Set Spline
This function permits calibration of the Scanner
position signals. This is required for the use of
spline curves in the
Line
scanning mode.
•
Click on
Set Spline
in the
Maintain
list.
−
The
Spline
window appears on the screen
(see Fig. 13).
•
You can set the required
Amplitude X
,
Amplitude Y
,
Offset X
and
Offset Y
values by
using the sliders.
•
Click the
Close
button to leave the window.
Fig. 12
Maintain list
Fig. 13
Spline window