NOTES ON DEVICE SAFETY
LSM 710 and LSM 780
Carl Zeiss
User Interface
Systems
32 M60-1-0025
e
02/2010
For NLO systems which are set up together with a Multiphoton laser, the beam tube connecting the
periscope and the scan head, has to be removed before the scan head can be dismounted. Perform the
following steps for the removal of the tube:
•
First pull out the tube from the periscope cover (Fig. 27/
5
) by sliding the inner tube into the outer tube
(Fig. 27/
4
).
•
Turn the tube assembly counter clockwise to unscrew the tube (Fig. 27/
4
) from the small black tube
(Fig. 27/
2
) which is fixed to the scan head (Fig. 27/
3
).
•
Now the tube can be removed and you can proceed as described above.
•
For rearranging the beam housing, first screw the tube onto the small black tube (Fig. 27/
2
) at the
back of the scan head (Fig. 27/
3
). It should not be fastened completely to allow the inner tube to fit
the side pins into the recess of the ring on the periscope (Fig. 27/
5
).
•
Open the shutter in the periscope using an Allen key (Fig. 26/
1
) and slide in the tube completely, this
will hold the shutter open.
Fig. 26
Shutter opening on the periscope.
Also applies for the large periscope
for upright microscopes.
Fig. 27
Tube connection from the periscope
housing (5) to the scan head (3).
Applies for upright and inverted
microscopes.