3
Advanced Maintenance
Replacing Components of the Analyzer
Thermo Scientific
ISQ 7000 Hardware Manual
171
7. Look inside the endcap and use a T10 Torxhead screwdriver to remove the two screws
holding the ion guide to the ion guide clamp.
8. Remove and replace the ion guide clamp.
9. Slide the ion guide and entrance endcap back onto the rods and exit endcap.
10. Reattach the ion guide and quadrupole to the analyzer tray. Make sure the quad and ion
guide are fully forward before tightening screws.
11. Reattach the screws on each side of the quadrupole rod endcaps
12. Attach the wire connecting the quadrupole entrance lens to the source interface board by
screwing in the terminal with a slotted screwdriver.
13. Once the components are correctly attached to the analyzer tray, follow the instructions
in
14. Restart your system by following the instructions in
Figure 142.
Detaching the Ion Guide from the Ion Guide Clamp
Figure 143.
Reattaching the Quadrupole and Ion Guide from the Analyzer Tray
Tip
You will get better signal if you gently push the quadrupole rod endcaps toward one
another while tightening the screws.
Screws
Ion Guide Clamp
Ion Guide
Screws
Exit Endcap
Entrance Endcap
Содержание ISQ 7000
Страница 1: ...ISQ 7000 Mass Spectrometers Hardware Manual 1R120617 0002 Revision C December 2018 ...
Страница 8: ......
Страница 174: ...2 Troubleshooting Contacting Technical Support 146 ISQ 7000 Hardware Manual Thermo Scientific ...
Страница 400: ...3 Advanced Maintenance Replacing Fans and Filters 372 ISQ 7000 Hardware Manual Thermo Scientific ...