Advanced Maintenance
Replacing the Calibration Gas Components
Thermo Scientific
ISQ 7000 Hardware Manual
359
20. Reinstall the RF board by aligning the connectors on the bottom of the board with the
connectors on the distribution board beneath it. It will drop into the slot when it is in the
right position.
21. Push the board down into the connectors.
22. Gently push the board toward the instrument, carefully aligning the board over the
feedthrough pins, until it stops.
23. Reattach the screw holding the RF board to the bracket.
Figure 404.
Reinstalling the RF Board
Figure 405.
Reattaching the RF Board and Bracket
RF Board
Chassis
Screw
Bracket