Performing Routine Maintenance
Replacing the Electron Multiplier
Thermo Scientific
ISQ 7000 Hardware Manual
103
15. Use a T20 Torxhead screwdriver to remove the four screws at each corner of the electron
multiplier plate.
Figure 100.
Disconnecting the Electron Multiplier Plate
Electron Multiplier
Plate Screws
CAUTION INSTRUMENT DAMAGE:
Use clean, lint-free gloves when touching
components inside the vacuum manifold.
ATTENTION DOMMAGES À L’INSTRUMENT
: utilisez des gants propres non pelucheux
lorsque vous touchez des composants à l’intérieur du collecteur à vide.