3
Advanced Maintenance
Replacing Components of the Analyzer
170
ISQ 7000 Hardware Manual
Thermo Scientific
Ion Guide Clamp
To replace the ion guide clamp:
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Remove the analyzer tray from the instrument by following the instructions in
“Removing the Analyzer Tray from the ISQ 7000 Instrument”
3. Remove the wire connecting the quadrupole entrance lens to the source interface board
by unscrewing the terminal with a slotted screwdriver.
4. Remove two screws on each side of the quadrupole rod endcaps.
5. Remove the entire quadrupole, as well as the ion guide, from the analyzer tray.
6. Slide the rods and exit quadrupole endcap off the ion guide and entrance endcap.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Figure 140.
Removing the Quadrupole and Ion Guide from the Analyzer Tray
Figure 141.
Separating the Ion Guide from the Quadrupole
Screws
Exit Endcap
Entrance Endcap