3
Advanced Maintenance
Replacing Components of the Front Door
286
ISQ 7000 Hardware Manual
Thermo Scientific
8. Attach new hinges and supports, as well as the hinge support bracket screw.
9. Reattach the nut and grounding washer.
10. Reattach the right side panel.
11. Close the front door of the instrument.
12. Restart your system by following the instructions in
Figure 301.
Reattaching the Nut and Grounding Washer
Grounding Washer
Nut