Preface
Hydrogen Safety Precautions
Thermo Scientific
ISQ 7000 Hardware Manual
xxi
Using Hydrogen with ISQ 7000 Mass Spectrometers
To use hydrogen with the ISQ 7000 instrument, you must always shut off the GC carrier gas
before venting or turning off the ISQ 7000 instrument. There are three hydrogen safety
screws on the ISQ 7000 instrument that
must
be in place. These are attached to your
instrument at the factory.
Make sure all the covers and panels of the ISQ 7000 instrument are firmly attached before
powering on the ISQ 7000 instrument. If you vented the system, make sure the vent valve is
tightly closed before powering on the system. Make sure all fittings, ferrules, and o-rings are
sealed prior to powering on the system.
Hydrogen Connection Guidelines
Use the following guidelines to safely connect hydrogen to your system:
•
Piping—
Hydrogen must be delivered to equipment using appropriate piping and be
done in such a way as to pose essentially no hazard to end-users. Piping systems for the
delivery of hydrogen should be designed and installed by a person qualified by specific
training and experience with hydrogen piping systems.
Figure 1.
Hydrogen Safety Screws on the ISQ 7000 Instrument
Right Top Cover
Screw
Front Panel
Screw
Left Top
Cover Screw