1
Performing Routine Maintenance
Cleaning the Ion Source Cartridge
Thermo Scientific
ISQ 7000 Hardware Manual
55
11. Pull the tool toward you and down the left side of the metal track on the barrel. As you
keep pulling it toward you, more and more of the barrel will be exposed.
12. Once you reach the end of the track (when the back engraved line at the end of the barrel
is exposed by the black handle), twist the handle to the left to lock it into place.
Figure 38.
Pulling the Tool Down the Left Track
Figure 39.
Locking the Tool in Place