Advanced Maintenance
Replacing Components of the Ion Source
200
ISQ 7000 Hardware Manual
Thermo Scientific
8. Reattach the thumbscrew-repeller spacers.
9. Reattach the source-repeller thumbscrews.
10. Reattach the manifold door and the four screws.
11. Restart your system by following the instructions in
Note
When reattaching the spacers, make sure they fit into the grooves on the
thumbscrews. Also, do not over-tighten the thumbscrews or it may crush the spacers.