3
Advanced Maintenance
Replacing Components of the Ion Source
Thermo Scientific
ISQ 7000 Hardware Manual
189
9. Remove the repeller plate and insulating spacers
10. Use a T10 Torxhead screwdriver to remove the two screws holding the gas mixing
chamber and tubing to the ion source block.
11. Use a T20 Torxhead screwdriver to remove the two screws connecting the lens/source
heater to the ion source block.
Figure 168.
Removing the Repeller Plate and Insulating Spacers
Figure 169.
Disconnecting the Lens/Source Heater
Repeller Plate
Insulating
Spacers
Screws
Lens/Source
Heater