3
Advanced Maintenance
Replacing Components of the Ion Source
190
ISQ 7000 Hardware Manual
Thermo Scientific
12. Remove and replace the ion source block.
13. Reattach the lens/source heater to the new ion source block and set them next to the rest
of the ion source.
14. Reattach the repeller plate and insulating spacers.
Figure 170.
Replacing the Ion Source Block
Figure 171.
Reattaching the Lens/Source Heater
Ion Source
Block
Screws
Lens/Source
Heater