13. Insert the ion guide and quadrupole onto the analyzer tray.
14. Reattach the screws on each side of the endcaps.
15. Attach the wire connecting the quadrupole entrance lens to the source interface board by
screwing in the terminal with a slotted screwdriver.
16. Follow the instructions in
17. Restart your system by following the instructions in
Figure 161.
Reattaching the Quadrupole and Ion Guide from the Analyzer Tray
Tip
You will get better signal if you gently push the endcaps toward one another while
tightening the screws.
Screws
Exit Endcap
Entrance Endcap