Upgrade Equipment
CI Reagent Gas Flow Module
Thermo Scientific
ISQ 7000 Hardware Manual
385
13. Look through the left side of the instrument and remove the screws holding the
dynode/multiplier power supply to the back and bottom of the instrument. Then remove
the power supply.
14. Attach the CI reagent gas controller cable to the distribution board.
Figure 441.
Removing the Dynode/Multiplier Power Supply
Figure 442.
Attaching the CI Reagent Gas Cable to Distribution Board
Bottom
Screws
Back
Screws
CI Reagent Gas
Flow Module
Connector