6. Set the rods on a clean work surface.
7. Use a T10 Torxhead screwdriver to remove the four screws around the rods. These screws
hold the wires in place. If the screws won’t come off the rods, you have to replace the
whole quadrupole.
8. Attach new wires to each side of the rods.
9. Slide the quadrupole rods back into the analyzer tray.
Figure 148.
Replacing the Quadrupole Wires
Screws
Screws
CAUTION INSTRUMENT DAMAGE:
Be careful when removing and inserting the screws
on the exit endcap. The screws are susceptible to galling in the rods.
ATTENTION DOMMAGES À L’INSTRUMEN
T : soyez prudent lorsque vous enlevez et
insérez les vis du bouchon de sortie. Ces vis peuvent se gripper dans les barres.