Performing Routine Maintenance
Cleaning the Ion Source Cartridge
72
ISQ 7000 Hardware Manual
Thermo Scientific
9. Place the locking ring on top of the repeller spring so that the repeller protrudes through
the center hole on the locking ring. The hooks on the sleeve fit between the larger gaps on
the locking ring.
At this point the instructions vary with the model of ISQ 7000 mass spectrometer. To
insert the ion source cartridge into the ISQ 7000 instrument with a VPI see
the Ion Source Cartridge in the ISQ 7000 with a VPI Instrument”
. To insert
the ion source cartridge into the ISQ 7000 instrument with no VPI, see
Ion Source Cartridge in the ISQ 7000 Instrument with no VPI”
Replacing the Ion Source Cartridge in the ISQ 7000 with a VPI Instrument
1. Hold the source holder and ion source cartridge in one hand and use the other hand to
hold the source exchange tool so that the barrel is facing toward the floor. The diagram on
the tool should be facing you.
Figure 66.
Inserting the Locking Ring into the Sleeve
Larger Gaps on
Locking Ring
IMPORTANT
If you try to attach the ion source cartridge to the tool without inverting it,
the components may fall, so make sure you are working over a table. Otherwise, you may
have to stop and clean the components again.