3
Advanced Maintenance
Replacing the Power Supplies
Thermo Scientific
ISQ 7000 Hardware Manual
349
Dynode and Multiplier Power Supply
To replace the dynode and multiplier power supply
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel and left side panel screws
from the interior front panel. Then slide the left and top cover panels toward the back of
the instrument and lift them off.
3. Use a T20 Torxhead screwdriver to remove the two screws holding the support bracket to
the middle of the instrument.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or the column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
Figure 389.
Removing the Controller Interface Board Support Bracket
Dynode/Multiplier
Power Supply
Support Bracket