Advanced Maintenance
Replacing Components of the Vacuum Interlock
Thermo Scientific
ISQ 7000 Hardware Manual
229
17. Remove the black axle bearing from the valve body. Replace the axle bearing, if necessary.
18. Push the valve axle into the side of the valve body. Replace the valve axle, if necessary.
19. Remove the brown axle seal from the valve body. Replace the axle seal, if necessary and
then insert it into the valve body.
Figure 229.
Replacing the Axle Bearing
Figure 230.
Replacing the Axle Seal
Axle Bearing
Clip
Safety
Interlock
Axle Seal