4
Upgrade Equipment
Turbomolecular Pumps
Thermo Scientific
ISQ 7000 Hardware Manual
403
10. Pull the adapter away from the turbo pump.
11. Use a 13mm wrench to loosen the four clamps at the top of the turbo pump.
12. Remove the two clamps that hold the turbo pump to the manifold on the transfer line
side.
13. Lift the turbo pump away from the remaining clamps.
14. Remove the turbo pump from the chassis. You may have to tilt the turbo pump to get it
out of the chassis.
Содержание ISQ 7000
Страница 1: ...ISQ 7000 Mass Spectrometers Hardware Manual 1R120617 0002 Revision C December 2018 ...
Страница 8: ......
Страница 174: ...2 Troubleshooting Contacting Technical Support 146 ISQ 7000 Hardware Manual Thermo Scientific ...
Страница 400: ...3 Advanced Maintenance Replacing Fans and Filters 372 ISQ 7000 Hardware Manual Thermo Scientific ...