1
Performing Routine Maintenance
Powering On the ISQ 7000 Mass Spectrometer
12
ISQ 7000 Hardware Manual
Thermo Scientific
7. Check the heater status by clicking the Status tab in the ISQ 7000 Dashboard. If the ion
source is not set to the right temperature, click the
Instrument Control
button and
change the temperature.
8. Check the status of the vacuum in the Status tab of the ISQ 7000 Dashboard. Within
10 minutes of powering on the ISQ 7000 mass spectrometer, the vacuum should read
OK. If it does not, see
9. Allow the ISQ 7000 instrument to stabilize for at least 30 minutes before running
samples.
Figure 4.
The ISQ 7000 Dashboard
Note
If the Status tab is not visible, click the arrow underneath the
Shut Down
button on
the dashboard to open it.
Note
You can run the instrument without stabilizing it, but you may have changes in the
masses and intensities as the system equilibrates at the final temperature.
Содержание ISQ 7000
Страница 1: ...ISQ 7000 Mass Spectrometers Hardware Manual 1R120617 0002 Revision C December 2018 ...
Страница 8: ......
Страница 174: ...2 Troubleshooting Contacting Technical Support 146 ISQ 7000 Hardware Manual Thermo Scientific ...
Страница 400: ...3 Advanced Maintenance Replacing Fans and Filters 372 ISQ 7000 Hardware Manual Thermo Scientific ...