3
Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
314
ISQ 7000 Hardware Manual
Thermo Scientific
13. Remove and replace the 20-pin feedthrough. If necessary, replace the o-ring and make
sure it is properly aligned and free of dust.
14. From the left side of the instrument, reattach the three screws around the 20-pin
feedthrough.
Figure 340.
Replacing the 20-Pin Feedthrough
Figure 341.
Reattaching the 20-Pin Feedthrough
O-ring
20-Pin Feedthrough
Screws