CI Reagent Gas Flow Module
382
ISQ 7000 Hardware Manual
Thermo Scientific
6. Look down through the top of the instrument and position the CI reagent gas flow
module on the interior side of the back panel.
7. Use a 1/8 to 1/16 in. reducing ferrule to attach the 1/16 in. stainless steel tubing to the
output connectors of the CI module.
8. On the exterior side of the back panel, use a T20 Torxhead screwdriver to attach the two
screws holding the CI reagent gas flow module to the instrument.
Figure 436.
Installing the CI Reagent Gas Flow Module
Figure 437.
Attaching the CI Gas Flow Module to the Instrument
CI Reagent Gas
Flow Module
CI Output
Connectors
Screws