Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
340
ISQ 7000 Hardware Manual
Thermo Scientific
15. Push the board down into the connectors.
16. Reconnect the Ethernet cable to the PC communication board.
17. Reinstall the RF board by aligning the connectors on the bottom of the board with the
connectors on the distribution board beneath it. The RF board will click into place when
its in the right position.
18. Gently push the board toward the instrument, carefully aligning the board over the
feedthrough pins, until it stops.
19. Reattach the screw holding the RF board to the bracket.
Figure 377.
Reinstalling the RF Board
Figure 378.
Reattaching the RF Board and Bracket
Note
This screw is required for good ground contact.
RF Board
Chassis
Screw
Bracket