3
Advanced Maintenance
Replacing the Calibration Gas Components
356
ISQ 7000 Hardware Manual
Thermo Scientific
7. Use your fingers to pry the bracket of the board away from the chassis and toward the left
side of the instrument. Lift the RF board out of the instrument and set it aside.
8. The lens driver board is the first board standing on end in the center of the instrument.
9. Look through the left side of the instrument and use a T20 Torxhead screwdriver to
loosen the screw holding the board to the chassis. Loosening the screw will separate the
board from the instrument. The screw is captured, so do not attempt to completely
remove the screw from the board and bracket.
Figure 400.
Locating the Lens Driver Board
Tip
The bracket has a white diagram describing the location of this screw. It is
approximately one inch from the top of the board directly under the white label.
Figure 401.
Removing the Lens Driver Board
Lens Driver Board
Bracket
Top Manifold
Cover
Lens Driver Board Screw