3
Advanced Maintenance
Replacing Components of the Electron Multiplier
266
ISQ 7000 Hardware Manual
Thermo Scientific
4. Use a T10 Torxhead screwdriver to remove the two screws holding the electrometer shield
in place.
5. Lift the electrometer shield off the instrument and set it aside.
6. Use a T10 Torxhead screwdriver to remove the four screws holding the electrometer
board in place.
Figure 282.
Removing the Electrometer Shield
Figure 283.
Removing the Electrometer Board
Electrometer Shield
Screws
Screws