Advanced Maintenance
Replacing the Calibration Gas Components
Thermo Scientific
ISQ 7000 Hardware Manual
365
Transfer Line
When replacing the transfer line, it is important to properly connect the nut to the transfer
line. Otherwise, you will get inconsistent or unreliable data because air or other contaminants
will get inside and interfere with the performance of the system.
To replace the transfer line and ferrule
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel screw from the interior
front panel. Then slide the top cover panel toward the back of the instrument and lift it
off.
3. Use a T20 Torxhead screwdriver to remove the screw at each corner of the right side
panel.
4. Slide the right side panel to the back of the instrument and lift it off.
IMPORTANT
Be sure to install the transfer line correctly or the ion source will not align
correctly.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering
these components.
Figure 414.
Removing the Right Side Panel Screws
Right Side Panel Screws